Product Details:
Minimum Order Quantity |
1 Unit |
Display Type |
TFT color LCD with LED back light |
Alarm |
Audio / Visual |
Probe Connector |
BNC Type |
Damping |
HI/LOW |
LCD Pixel Density |
320 x R.G.B. (H) x 240 (V |
DIGISCAN DS-333 is built tough and easy to use. Full size access to all the functions. The large, high resolution display has a wide viewing angle and good key board ensures direct visibility in any light conditions.
The TFT colored display has the ability to display small echoes which are very relevant to demanding UT operators, it does not miss any information. The DAC is offered in accordance to demanding customers needs.
DIGISCAN DS-333 is a complete instrument suitable for most applications where Ultrasonic Testing is carried out. It's simplicity to use, versatile features and ability to display high fidelity Ultrasonic signals, makes it an ideal choice for Axle Testing.
Features:
- Operation with Li-Ion Battery
- 3 Kg weight with batteries
- Receiver Bandwidth range from 1MHz to 6MHZ
- Selectable transducer modes: single, dual and angle beam
- Dual gate for accurate thickness measurement
- Built in trigonometric function for precise flaw detection
- Internal triggering
- Gated zoom facility to view expanded portion of the defect
- Selectable logic for both gates
- Background reference echo pattern for easy comparison of the defect
- 10000 ‘A’ scan memories
- 100 calibration sets
- USB interface for data logging
LCD Display:
- Display: TFT color LCD with LED back light, 300 X 200 pixel, 120mm (W) X 90mm (H) Pixel density 320x R.G.B(H) x 240(V)
- Display Freeze: Current display freezed
- Active Freeze: Display accumulation, active freeze echo to be saved
- Reference Echo Pattern: Reference pattern can be displayed at the background from selected location in memory
- Display Pattern: Unfilled or Filled
PULSER / RECEIVER:
- Receiver Bandwidth: 1MHZ to 6MHZ by broad band amplifier
- Gain: Total gain 120dB, 0–80dB operatable, 0.1,0.5,1,2,6,12dB selectable
- Rejection: Linear (0 – 99%)
- Operating Mode: Single Probe T+R, double probe T/R
- Test Mode: Pulse reflection or pulse Transmission
- Energy: NOR/LOW selectable
- Probe Zero: 99.99us in step of 0.1mm
- Vertical Linearity: Within ±3% on all ranges
- Probe Connector: BNC type
- Echo Display: Full wave rectified
Time Base:
- Range: 5mm min. –5000mm. Max (@5890M/sec)
- Delay: 0-3000mm continuous variable in step of 0.1mm, 0.01mm/step for 10-100mm, 0.1mm/step for 100mm–1Mtrs, 1mm/step for 1Mtr–5Mtr
- Monitor Gate: 2 Gates - Width 1/10 screen to full screen, Height 1/10 screen to 99%, Gate start measured from '0' div & ± logic selectable
General:
Memory:
- Minimum 10000 Sets data set-ups along with ‘A’ scans with calibration Parameters, trigonometric values and Label
- 100 calibration parameter sets
- Recalling the cal-sets, Frame set and labels as and when required
Interface / Outputs:
- PC connectivity via USB-A, with Export function opening in MS Word / PDF
- Pen Drive Connectivity via USB-B.
- Direct Printer Connectivity-Dot matrix Printer-DB9 Connector
PC Software:
EECOWIN PC interface software with Suitable port & transfer software. Can upload “FRAME” from saved set to PC through the USB port and through Pen drive.
Power Source:
Li-ion battery 3.7V max, 20800mAH
Rechargeable battery model BAT324l
-Operating time 8Hrs (min) with battery
-Battery charger cum mains adaptor 100-240V, 50Hz/60Hz, Model
-AD5V5 Full charge in 8hrs
-External charger model CGR322 with automatic cut-off facility to protect battery from over charging