Technical Specifications |
Test Range |
2.5mm to 10 meter (0.100 in to 400 in) (longitudinal waves in steel). Fine mode is adjustable in minimum step of 0.01mm (0.001in). |
Velocity |
1000 met. /sec to 15000 met. /sec (40 in/millisecond to 600 in/millisecond). In Hot key mode it has pre-set values. Fine mode it is adjustable in step of 1met/sec.(1in/millisecond). |
Delay |
Variable from -10 to 2000mm (-0.3 to 30 inches). |
Gain |
100 dB calibrated gain adjustable in 0.1, 0.5, 1, 2, 6 or 12 dB step. |
Rejection |
0 to 100% FSH with Led indicator. |
Rectification |
Full-wave rectified, Half wave – ve, Half wave + ve, and RF mode. |
Frequency |
It has tuned amplifier with four bands: a) 0.2MHz to 1MHz b) 0.5MHz to 4MHz c) 0.8MHz to 8MHz d) 2MHz to 20MHz |
Linearity Deviation |
Vertical ±3% Horizontal: ±0.5%. |
Test Modes |
Pulse echo and Transmit/Receive. |
Transmitter |
Transmission pulse negative spikes. (Pulse rise time<10ns) and with Selectable high (300 Vp) or low (250 Vp) power. |
Damping |
Damping high/low is selectable. (High= 45Ώ, Low=345Ώ). |
Connectors |
BNC or LEMO size-1 factory optional |
Freeze/ Peak Freeze |
A-scan Freeze, Peak freeze, echo dynamic available. In peak freeze it holds peak amplitude pattern which is useful for angle beam probing to to locate peak signal. |
Monitor |
Dual gate with threshold adjustable in 1% of the screen height with +ve/-ve, Gate expand Interface trigger modes. |
Gate Expand |
Expands range to width of the gate. |
A-scan Memory |
500 trace patterns can be stored which can be recalled, printed, or transferred to PC via USB (with NOTE/DETAIL). Unlimited No of A-scan can be directly stored in USB disk with auto file naming. |
B-scan Memory |
50 B-scan can be stored (with NOTE/DETAIL) which can be recalled, or transferred to PC via USB. Unlimited no of B-scan can be directly stored in USB disk with auto file naming. |
Calibration Set-up |
50 different calibration set-ups can be stored and recalled. |
T-LOG |
20000 readings can be stored in 20 different files. Five different types of file templates for file creation. Stored readings can be transferred to PC via USB. |
DAC |
Dynamic DAC curve can be digitally plotted (smooth parabolic curve) on screen with selectable additional offset curves from 0 to 14 dB in 0.1dB selectable steps. DAC curve can be set as a flow monitor gate. DAC can be plotted using minimum 2 to maximum 10 points. |
TCG |
After plotting DAC , TCG( Time corrected Gain) can be activat
|